Overview of oxygen connector types
Contributors: Michael Lipnick, MD, Sky Vanderburg, Cornelius Sendagire, MMed, Robert Neighbour
Date last updated: May 19, 2023
Are you finding it difficult to figure out what type of oxygen connector you have or need? You are not alone. Unfortunately there is no universal standard for oxygen connectors. This article will review the most common types of connectors used for connecting oxygen supplies to oxygen delivery devices. We will break down the discussion into:
- Low pressure (<30 psi, <2 bar)
- Intermediate (~50 psi, 3.5bar)
- High pressure (>2000 psi, > 137 bar)
Always consult the manufacturer’s specifications and recommendations prior to connecting any oxygen equipment. Improper use or ‘DIY’ approaches can be dangerous. In most countries, parts must comply with national regulatory body standards. Check to see what regulatory bodies are followed in your country. There are multiple agencies dedicated to medical gases includingEIGA (Europe); AIGA (Singapore); ANZIGA (Australia/New Zealand); CGA (United States of America); JIMGA (Japan); and SACGA (South Africa).
An overview of oxygen delivery devices and oxygen sources can be found elsewhere in the O2 Encyclopedia.
Low pressure oxygen connectors
Low pressure is not strictly defined when discussing oxygen delivery. Here we use the term to describe oxygen delivery from sources with <2 bar (~30psi) at the takeoff from the source. Most low pressure oxygen delivery sources are far lower in pressure.
Flowmeters
Thorpe tube ('rotatmeter') flowmeter
- O2 Flows: depends on prespecified size (0-200mL, 0-1 LPM, 0-5 LPM, 0-15 LPM, 0-25 LPM, 0-80 LPM)
- O2 Source: Must be ~3.45 bar (~50 psi) – either outlet from regulator on cylinder or pipe terminal (wall-unit)
- Age consideration: all ages, though for pediatric and especially neonatal patients may require precisely controlled, lower flow devices
- Merits: Easy to use, common, displays/measures flow, if pressure compensated then accurate despite back pressure (e.g. kinked or occluded delivery device)
- Drawbacks: easily breakable, must be used upright (gravity impacts accuracy), if pressure uncompensated then may display inaccurate flow if back pressure; requires pressure reducing valve if O2 source >3.45 bar/50 psi
The most common type of low pressure oxygen connector is the outlet from a Thorpe tube (‘rotatmeter’) flowmeter (for more info on other types of flowmeters in use the slider at the top left to see a more ‘advanced’ version of this article). Flowmeters help not only control the rate of oxygen flow to a patient but they also measure it – providing feedback on consumption to the clinician.
Flowmeters come in many different preset flow delivery ranges (0-200mL, 0-1 LPM, 0-5 LPM, 0-15 LPM, 0-25 LPM, 0-80 LPM and more). It is important to select the flowmeter range based on the patient population (e.g. neonates may require very precise control < 1LPM) and the delivery devices used (e.g. adult high flow nasal cannula system requires 60LPM).
Flowmeters may be connected in different ways to the oxygen source. A Thorpe flowmeter must be connected to 3.45 bar (50 psi) oxygen pressure to function properly. This means a Thorpe flowmeter can be connected directly to the outlet from a regulator (~55 psi) from a cylinder, or the flowmeter may be connected to a terminal wall unit of a piped system (up to 55 psi). By the time oxygen leaves the flowmeter, it is stepped down to lower pressure, and thus we are including this in the low pressure section (more on regulators later).
Flowmeters may also be built into low pressure sources like an oxygen concentrator (so there may be no need to procure a separate flowmeter). Most portable bedside oxygen concentrators have a built-in flow meter to adjust flow to the patients. Flow and pressure vary by brand/model though most commonly are in the 0-10 LPM range.
Flowmeters (which are not already built into the cylinder or oxygen concentrator) have an input connector and an output connector. The input connector of the flowmeter can be one of multiple types, though commonly is a diameter index safety system (DISS) connector (other types are discussed in high pressure connector section below).
The output connector is most often a standard threaded cylindrical connector which accepts a barbed, tapered adapter discussed in the next section. A humidifier may also be connected directly to the flowmeter as discussed in the ‘humidification systems’ section below.
Flowmeters help not only control the rate of oxygen flow to a patient but they also measure it – providing feedback on consumption to the clinician.
There are three common types of gas flowmeter:
- Thorpe tube (‘rotameter’) – variable orifice; comes in either back pressure compensated or uncompensated type. Back pressure compensated – the float is between the valve and the oxygen source, so if pressure is applied distal to the tube (e.g. a kinked tube or obstructed outlet) then the flowmeter will still function accurately (i.e. the float will drop down until the obstruction is relieved. Requires separate pressure-reducing valve if used on higher than 3.5 bar/50psi source; works only in vertical position (i.e. affected by gravity), fragile if dropped, if uncompensated then can display erroneously high flow rate
- Bourdon gauge – fixed orifice, integrated pressure reducing valve, accepts variable inlet pressure, used only with cylinders, unaffected by gravity, not back pressure compensated; with increased flow the indicated flow reading becomes inaccurate
- Dial/click meter – fixed orifice, calibrated to deliver specific flow in increments; has integrated pressure-reducing valve; commonly built into small oxygen cylinders; accepts variable inlet pressure, unaffected by gravity, does not measure flow, expensive
Thorpe flowmeter configurations
How to read a Thorpe tube flowmeter with ball or bobbin mechanism (4 LPM)
Tapered, Barbed Connector
The tapered, barbed connector is a commonly encountered component for connecting oxygen delivery devices to flowmeters (e.g. flowmeter on a portable oxygen concentrator, cylinder or wall connected flow meter). This connectors is sometimes referred to as a ‘Christmas (fir) tree connector’ because of its shape and sometimes green color.
This connector comes in plastic or metal and simply screws on by hand (and can be damaged by over tightening).
Low pressure smooth bore oxygen tubing is connector to the barbed end of this adapter, as discussed in the next section.
CAUTION: this connection is not secured and accidental disconnect is common (especially if not applied with adequate force or if pressure is high)
Barbed oxygen connector ('Christmas tree' adapter)
Smooth bore oxygen tubing
Smooth bore oxygen tubing is an essential piece of oxygen delivery equipment. It is designed to be used with low pressure oxygen sources. This tubing comes in different colors (green or clear) and lengths and is often integrated into the delivery devices (e.g. nasal cannula or facemask). To connect smooth bore oxygen tubing to the oxygen source, the end of the tubing with a tapered opening is simply pushed over the tapered, barbed connector of the source by hand.
Two-sided, tapered, barbed plastic adapters may be obtained to connect smooth bore oxygen tubing extensions when needed.
Oxygen tubing must adhere to many specifications to ensure safety – these include transparency, strength, anti-kink design and other features.
(Use the slider at the top left of this page to see more advanced information on oxygen tubing)
Examples of smooth bore oxygen tubing
Flow splitting devices
A number of devices and techniques are used for splitting one oxygen source to multiple delivery devices. Not all are recommended and some can be dangerous for both patients and providers. Always consult the manufacturer’s specifications.
Oxygen connectors are often improvised in resource-constrained settings (see image below), but this practice can be very dangerous.
Devices that are often used for splitting include:
- Flowmeter stands (aka flowmeter stations or assemblies) – these boxes allow input from one oxygen source and then split that source into multiple outlets, each controlled by an independent flowmeter. These can be costly and require clear labeling of flowmeter-bed to avoid inadvertent titration for the wrong patient. These can be connected to an oxygen concentrator, though max total flow will be limited by the output of the oxygen concentrator.
- Dual flowmeters – as the name implies, this is a flowmeter with input from one oxygen source but output to two flowmeters. These require a high pressure oxygen source.
- Plastic flow splitter – this may be as simple as a y-connector. Setups using plastic flow splitters may result in unequal distribution of oxygen to patients due to limited ability to measure and control flow
Flow splitting via flowmeter station and y-splitters
Improvised oxygen delivery setups
Humidification systems
There are multiple types of humidification systems and many can be placed in a low pressure oxygen delivery setup. The most common is the non-heated bubble humidifier. These simple devices come as either single use or reusable units and connect directly to a flowmeter. These devices require sterile or distilled water and are generally not efficient as compared to heated humidification systems (for full discussion, see Respiratory care humidifications systems)
Bubble humidifiers reduce the dryness of the oxygen supplied by bubbling the gas through water at room temperature. They must be filled with clean water at least once a day (distilled water, sterile or tap water that has been boiled and cooled) and the water level should be checked twice daily. The humidifier equipment must be washed and disinfected regularly to prevent bacterial colonization.
Humidification is likely indicated with higher O2 flow rates through nasal prongs (>1 L/min for a neonate, >2 L/min for an infant, >4 L/min for an older child or adult) or with devices that bypass the humidifying nasal passages (e.g. nasopharyngeal catheter).
Bubble humidifier
- O2 Flows: utilized with higher O2 flow rates (>1 L/min for a neonate, >2 L/min for an infant, >4 L/min for an older child or adult) through nasal prongs, or with devices that bypass the humidifying nasal passage (e.g. nasopharyngeal catheter)
- O2 Source: Thorpe tube flowmeter
- Age consideration: all ages, though for pediatric and especially neonatal patients may require precisely controlled, lower flow devices
- Merits: Easy to use, common, no electricity required
- Drawbacks: require sterile or distilled water (or boiled and cooled); if used for too long or not properly cleaned (reusable units) can become source for infection; no heating provided and not efficient at delivering humidification (especially if temperature less than 30 degrees)
Intermediate pressure oxygen connectors
There is not standard definition for intermediate pressure oxygen. Here we define ‘intermediate pressure’ as oxygen pressure at ~3.45 bar (50 psi). Many oxygen delivery devices require this pressure for safe function. In other words they will not function with unregulated, high pressure directly from a cylinder, and would also not work with low pressure from a portable oxygen concentrator. Of note, some ventilators, CPAP/BIPAP/NIPPV machines, and high flow nasal cannula have built in turbines or compressors that may allow them to function without an intermediate pressure oxygen source.
Hose & pipe terminal connectors
There are more than a dozen ‘standards’ for high pressure oxygen hose or wall fitting around the world. Unfortunately, many countries and regions have created their own standards which are not cross compatible. In some regions, especially in low and middle-income countries, it is common to encounter multiple different types of connectors. In Canada and the US, new construction uses the CGA DISS connector, though multiple ‘quick connect’ styles are commonly found. These quick connect devices may provide convenience but at a potential cost for durability and reliability.
Here we summarize the most common 18 high pressure hose and piped terminal oxygen connector – DISS, NIST, Puritan Bennett, Japanese Style, British Standard, German DIN, AGA Swedish, Italian UNIFOR, Ohmeda quick connect, DISS, Chemetron, Oxyequip, French AFNOR, Australian SIS, Schrader and more.
High pressure oxygen hose and terminal connector types
Reinforced hose
Connecting to an intermediate pressure (3.45 bar/50 psi) source requires specializes oxygen tubing to ensure safety. The smooth bore oxygen tubing described above is not sufficient – the tubing is not able to maintain integrity and would not remain securely connected to the source during intermediate pressure.
Rubber tubing that is reinforced (often braided polyester yarn coated in rubber) and intended for medical use is required. Color coding of such tubing and end connectors will vary by region. The working pressure of the hose may vary by the gas it was intended to be used with – ie make sure you are using the correct type of gas hose. Always check with local manufacturer’s and standards.
Intermediate pressure reinforced oxygen hose
Oxygen ecosystem: from source to bedside
High pressure oxygen connectors
High pressure (in the context of clinical oxygen delivery systems) is most commonly defined as oxygen pressure at 100-200 bar (>2000 psi) or higher. This pressure is unsafe to deliver to patients and must by regulated to a lower pressure before connecting to medical equipment or oxygen delivery devices.
Cylinder regulators
Cylinder regulators serve to modulate pressure from a high pressure (100-200 bar) oxygen source to provide a reduced and stabilized intermediate outlet pressure (usually 3-6 bar). Importantly, properly functioning regulators enable the outlet pressure/flow to remain near constant regardless of fluctuations in pressure of the oxygen cylinder. Regulators which provide a stabilized flow output (otherwise known as flow gauges) have a flowmeter integrated into the same mechanical fitting. It is also possible to connect a flowmeter to the outlet port of a regulator stabilizing pressure to supply oxygen directly to patients (see image below).
Regulators which provide a stabilized pressure are used in conjunction with equipment (i.e., anesthesia machines, ventilators, high flow nasal oxygen delivery devices, etc.), while regulators which provide a stabilized outlet flow can be used to supply oxygen directly to patients.
There are multiple types of regulators, including:
- Single gauge regulator – has a set pressure output and displays pressure of the source content. Can be connected directly to a patient via a flowmeter
- Dual gauge regulator – with one gauge (‘inlet gauge’) showing the pressure of the source content and the second gauge showing the reduced outlet pressure which can be controlled with a knob or screw.
It is important to ensure a regulator outlet pressure is in the desired range and fits the cylinder (i.e. gas type).
Flowmeter attached to pressure gauge for oxygen delivery to patients
Inner Components of a Pressure Regulator
Bourdon gauge
Bourdon gauge
Bourdon gauges measure the pressure of a gas contained in a compartment (e.g. cylinder). They are the most commonly used pressure gauges in medical gas equipment. Of note, Bourdon gauges display ‘gauge pressure’ which is not the same as ‘absolute pressure’ which includes atmospheric pressure, so Bourdon gauges function similarly at any altitude.
Cylinder valve types
Dial (click) meter
Dial/click meter
Dial (click) meters are calibrated to deliver a set flow at a constant pressure (3.45 bar, 50 psig). The flow is selected via an adjustable dial which displays the flow in a small window. Dial (click) meters come in various ranges of flow depending on the patient care need. Importantly, dial (click) meters are only accurate at their rated pressure.
Integral valves
This type of valve has an integrated pressure regulator and flowmeter and are found more commonly on smaller cylinders.
Pin index/Yoke valve (CGA870; ISO 407)
Pin index/Yoke valve
Pin-index/Yoke valve, side spindle
Pin index/yoke valves utilize the Pin Index Safety System (PISS) which is designed to ensure the correct connection via a unique pin arrangement by gas type. For example, pin index/yoke valves for cylinders containing oxygen have the specific pin arrangement displayed above (*click on Pin index/Yoke valve image to display).
Bull nose valve (CGA580; BS 341)
Bullnose valve
Bullnose connector
In the US and Canada, CGA 580 is considered standard for industrial use while CGA 540 and 810 are considered standard for medical use.
Handwheel valve (CGA540)
Handwheel valve
Handwheel valve, side outlet
Handwheel valves
Oxygen connector leaks
Leaks in oxygen infrastructure and delivery devices are common. Common causes to include:
- User error – e.g. using excess flow
- Not turning off devices that are not in use (e.g. anesthesia machines or flowmeters)
- Incorrectly secured fitting from cylinders
- Leaks in high or low pressure tubing due to lack of adequate maintenance
There are many ways to conserve oxygen, including through regular inspection and routine maintenance.
Oxygen connector leaks
Oxygen connector leaks
References
- WHO-UNICEF technical specifications and guidance for oxygen therapy devices. Geneva: World Health Organization and the United Nations Children’s Fund (UNICEF), 2019 (WHO medical device technical series). License: CC BY-NC-SA 3.0 IGO.
- Frank’s Hospital Workshop. Frank Weithöner, 2022. frankshospitalworkshop.com